XL 640 is a fully automated biochemistry analyzer with a throughput of 640 tests per hour. It is a dependable, powerful and robust analyzer to keep up with the heavy demands of everyday. XL 640 is designed to operate smoothly over the long haul and to deliver consistent quality report. XL 640 provides innovative solutions that meet the high productivity demands of laboratories having heavy testing pressure. It is equipped with excellent features which yield higher productivity, increased efficiency and greater cost savings for the laboratory. It has got the continuous loading facility for samples which make XL 640 the most dependable partner.
Product Applications:
System Type: Automated, Discrete, Random Access, Patient Prioritized Clinical Chemistry Analyzer
Throughput: 400 photometric tests/hour & 640 tests/hour with ISE (*optional ISE with Na, K, Cl, Li)
Analytical Methods: 1-Point, 2-Point, Rate-A, Rate-B, Direct Potentiometry(optional)
Barcode Reader: Both for Samples & Reagents (Optional)
Reactions Tray: 72 hard glass cuvettes
Reaction Mixing: Stirrer with variable speed
Reading Volume: 180 µl
On-Board Laundry: 6 stage cleaning, 2 stage drying
Photometer: Multi-wavelength diffraction grating with 12 wavelengths (340, 376, 415, 450, 480, 505, 546, 570, 600, 660, 700 & 750 nm)
O.D. Range: 0 – 3.0
Light Source: Halogen Lamp
Detector: Silicon Photodiodes
Water Consumption: < 14 liters/hour
Programmable Parameters: Default system pack parameters + upto 99 user defined parameters, Unlimited profiles and unlimited calculation items
Quality Control: Q.C. plot data with Q.C. Rules. Provision for lab mean. Twin plot.
Caliberation: K- factor, Linear (1,2 point and multipoint) 4P and 5P Logit-log, Cubic spline,exponential, polynomial. On-board serial dilution for calibrator
Power Requirement:
AC 220 V ± 10 %, 50 Hz or
AC 110 V ± 10 %, 60 Hz
Power Consumption: 800 VA
Dimensions: Approximately 897 mm (w) x 655 mm (D) x 1170 mm (H)
Weight: Approximately 200 Kg
Sample: Serum, Plasma, Urine, CSF, Whole blood, other
Sample Unit: upto 82 positions for samples/STAT/calibrator/controls
Sample Pipetting: 2-70 µl (adjustable in 0.1 µl steps) for biochemistry,70 µl fixed for ISE capacitance probe with liquid level sensing and vertical obstruction detection , clot detection and serum indices.
Auto Rerun: Repeat with same, increade and decresaed volume (upto 1:150)
Sample Tubes/Cups: Primary tubes of 5 ml, 7ml and 10 ml and sample cups
Reagent Tray: 56 positions for reagents with on-board cooling
Reagent Pipetting:
R1: 60 – 300 µl (Adjustable in 1 µl step)
R2: 0 or 10 – 300 l (Adjustable in 1 µl step)
Capacitance probe with level sensing and vertical obstruction detection
Analyzer-PC: USB
PC- HOST Computer: Bidirectional TCP/IP and RS – 232
PC- Printer: USB
Operating System: Windows based
Database: Unlimited Results